Installation guide

UltiMate
®
3000 Viper™ Capillary Kits for Intelligent LC (LCi) System Solutions
Quick Installation Guides
Page 28 Rev. 02 January 2010
© 2010 Dionex
Table 10: Detailed description of capillary connections of the Parallel LC setup (dual-stack)
Description of Connections Capillary Dimensions (ID x L)
PumpLeft outlet TCC 6-port valve left, port 6 0.18 mm/0.007” x 550 mm
PumpRight outlet TCC 6-port valve left, port 4 0.18 mm/0.007” x 350 mm
TCC 6-port valve left, port 5 WPS sampler valve, port 5 RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
WPS sampler valve, port 4 TCC 6-port valve left,
port 2
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 3 Column 1 inlet RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 1 Column 2 inlet RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
Column 1 outlet Detector 1 inlet RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Column 2 outlet Detector 2 inlet RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Note: In case of a single-stack setup of the Parallel LC system, the TCC 6-port valve
left turns into a TCC 6-port valve right. The capillary lengths have to be
adapted as outlined in Table 11.
Table 11: Detailed description of capillary connections of the Parallel LC setup (single-stack)
Description of Connections Capillary Dimensions (ID x L)
PumpLeft outlet TCC 6-port valve left, port 6 0.18 mm/0.007” x 950 mm
PumpRight outlet TCC 6-port valve left, port 4 0.18 mm/0.007x 950 mm
TCC 6-port valve left, port 5 WPS sampler valve, port 5 RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm
WPS sampler valve, port 4 TCC 6-port valve left,
port 2
RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm
TCC 6-port valve left, port 3 Column 1 inlet RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 1 Column 2 inlet RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
Column 1 outlet – Detector 1 inlet RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Column 2 outlet Detector 2 inlet RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm