User's Manual

Rev.1.0
Page 145 of 151 VW40-153-006
VIVIX-S 1717N User Manual
9.4 Guidance and Manufacturers Declaration for EMC
This device has been tested for EMI/EMC compliance, but interference can still occur in an
electromagnetically noisy location. Attempt to maintain a suitable distance between electrical
devices to prevent malfunction.
9.4.1 Electromagnetic Emissions
The Equipment Under Test (EUT) is intended for use in the electromagnetic environment specified below.
The customer or user of the EUT should assure that it is used in such an environment.
Immunity test
Compliance
Electromagnetic Environment
RF Emissions
(CISPR 11)
Group 1
The EUT uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
(CISPR 11)
Class B
The EUT is suitable for use in all establishments other than domestic
and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
(IEC 61000-3-2)
Class A
Voltage fluctuations/
Flicker emissions
(IEC 61000-3-3)
Complies
9.4.2 Electromagnetic Immunity
The VIVIX-S 1717N system is intended for using in the electromagnetic environment specified below. The
user of this system should assure that it is used in the following environment.
Electrostatic Discharge (ESD) IEC 61000-4-2
Item
Description
Immunity test
Electrostatic discharge (ESD) IEC 61000-4-2
IEC 60601 test condition
Contact ±6
Air ±8
Compliance Level
Contact ±6
Air ±8
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity
should be at least 30%.