User's Manual

Rev.1.0
Page 130 of 151 VW40-153-006
VIVIX-S 1717N User Manual
Solutions
1
Check the connection between AC power cable and SCU Basic.
2
Check the connection between DC power cable and SCU mini.
3
Turn off the power switch and turn on again and then check the fan or status of back side.
4
Replace another SCU and check the result of it
5
Replace the fuse of standard SCU. (refer to <7.6 Replacing the Fuse of SCU (SCU Basic only)>)
6
Replace corresponding devices.
8.1.4 Communication Test is failed
Symptom
Transmission error is occurred, failure of communication test is occurred.
Expected Causes
Network connection problem
Network setting problem
PC environment setting problem
Wireless environment environment problem
Devices error
Solutions
1
Check the connection of network cable between Workstation and SCU.
2
Check if the accurate network cable is used or not. (CAT 5E or 6)
3
Set the network information of Workstation, SCU and detector again.
4
Check whole workstation environment again such as firewall setting and release the power save mode.
5
Check surrounding wireless communication environment.
6
Boot up detector and SCU again by processing initialization. (Refer to <7.5 Product Initialization>.)
7
Replace other SCU and Detectors and check the result of it
8
Replace corresponding devices.
8.1.5 The Active LED and Data LED of the Detector are blinking
Symtom
The active LED and data LED is blinking when power LED is ON.
Expected Causes
Detector registeration error
Data transmission error
Solutions
1
Turn on SCU again
2
Check the network cable connection
3
Check the workstation environment and network information again
4
Check if the surrounding wireless communication is good