VIVIX-S 1717N User Manual
8.1 Trobleshooting
8.1.1 Troubleshooting Guide
When you encounter problems while using the equipment, search for the table below for the problem or
error messages and try the solutions.If the problem persists, turn off the detector and consult your sales
representative or a distributor. Please refer to the details of the following symptoms or error messages.
Troubleshooting must be performed by service engineer who is authorized by Vieworks. If
an unqualified person performs troubleshooting on the system resulting in damaging the
detector, software or hardware, then the Vieworks or its representative is not responsible
for the detector repair regardless of remain warranty. For more detailed information, refer
to <10.1 Service Information> and <10.2 Warranty>.
8.1.2 Failed to Turn the Detector On
Symptom
Failed to turn on the power of the detector.
Expected Causes
Not installing a battery pack properly.
Dead battery pack
Battery pack or the detector is broken down.
Solutions
1
Install battery pack
2
Charge battery pack
3
Check the result after getting rid of the battery pack and connecting the tether cable.
4
Replace other battery packs and check the result of it.
5
Replace another detector and check the result of it.
6
Replace corresponding devices.
8.1.3 The Power Switch of SCU or Status LED is not worked
Symptom
The power switch of Basic SCU or SCU mini is not working.
The status LED of SCU is not responding.
Expected Causes
Power cable is broken down.
Errors in the fuse
Internal circuit is broken down.