7.3 Self Diagnosis
You can perform self-diagnosis of the detector and SCU from the Self Diagnosis tab of the Diagnosis
dialogue in VIVIX Setup.
1
Choose a desired item to diagnose.
2
Click Execute Test button located at the bottom of the Test Case window.
3
Check the progress and result of diagnosis for each item in the Result window.
4
Click Save to File to save the diagnosis result as a file when the self diagnosis is completed.
7.3.2 Self-Diagnosis Items of Detector and Measures
Voltage
Item
Form
Expected problem
Measures
DDC
Decision
Defective tether interface cable
Change a tether cable.
Poor power supply to the wired operation mode.
Contact a service engineer.
MAIN
Decision
Poor power supply to the processor.
Contact a service engineer.
SIGNAL
Decision
Poor power supply to FPGA.
Contact a service engineer.