User's Manual
Rev.1.0
Page 116 of 151 VW40-153-006
VIVIX-S 1717N User Manual
7.1.2 Battery Pack Diagnosis
Install the battery pack to the detector and check the voltage and remaining amount of
the battery pack. Furthermore, always check the remaining amount of the battery pack
during use of the detector. If performance of the battery pack has some problems, consult
the sales representative in Vieworks or a relevant engineer.
The battery pack belongs to consumables which performance will be decreased as time
passed. Make sure to check the battery life during usage.
Check remaining amout of battery pack
The battety remains is noticed as a level or percentage (%).
Check the battery status from the Information tab of the Diagnosis dialogue in VIVIX Setup program.
It is also available to check the battery remains from the LED on the side of detector.
You can check the battery remains from VXvue (Vieworks Viewer) or VIVIX SDK.
Item
Description
Equipped
The install state of a battery pack. (Equipped / Not Equipped)
Charging State
The charging state of a batter pack. (Charging / Not Charging)
Remain
Battery remains
Voltage
Battery voltage
Display of Bettery Remains
Level
Value
Battery Remains
LED Display
Full
5
81% ~ 100%
5
th
level
Half Quarter
4
61% ~ 80%
4
th
level
Half
3
51% ~ 60%
3
rd
level
Quarter
2
31% ~ 50%
2
nd
level
Low
1
11% ~ 30%
1
st
level
1% ~ 10%
Blink
Unknown
0
Off