VIVIX-S 1717N User Manual
7.1 Diagnosis
7.1.1 Image Diagnosis
Check the image quality through Diagnosis tools after installing the detector or before
usage. If the problems with regard to products or image occur during diagnosis, try to do
a calibration again. If the problems are not solved, consult the sales representative in
Vieworks or a service engineer.
You can acquire and review an image from the Image window in VIVIX Setup program.
You can review images by acquiring them through real exposure or getting Dark image by clicking Get
Normal Image button.
The number of images, pixel value and ROI value will be displayed.
The effective area or whole area of an image can be checked.
It is also available to check the image by changing its direction.
Save the reviewed image as a raw one to analyze.
You can acquire an image either applying Offset / Gain data or not.
[TBD] Screen Shot
Diagnostic Functions
Item
Description
Pan
Press and drag a mouse button to move the image to the desired position.
Zoom
Press and hold a mouse button to move the image upwards or downwards for
expanding / reducing the image.
W/L
Press and hold a mouse button to move the image to up/down/right/left for
adjusting its window level. This function can be used without clicking W/L button, but
with the right mouse button basically.
Fit
Sets the image to the center, which was moved by using Pan function.
Statistic
Clicks and drags the left mouse button to set arbitrary area. The coordinate, min/max
value, average and standard deviations are displayed on Pickup ROI at the left side
of image.
ROI
Clicks and drags the left mouse button to set window level automatically on a basis
of min/max value of the configured area.
Zoom
Expands the image
X1
2 times
X2
4 times
X4
16 times
Offset
Applicability of Offset data