VIVIX-S 1717N User Manual
6.5.4 Auto Defect Correction
While operating the detector, defect pixels may appear on the image. In this case, use the
defect auto correction to calibrate the defect pixels.
Be sure to check the followings before acquiring the FLAT image.
Preheat the detector for 30 minutes or more.
The recommend SID is 150 (distance between X-ray tube and detector)
Open the collimator of X-ray tube completely.
Align the center of the detector with the center of collimator.
Keep everything away from the detector surface.
Adjust the x-ray dose to make the pixel value from 900 ~ 1100.
Check if the calibration data (Offset, Defect Map, Gain) is registered normally.
1
Make an exposure of FLAT images from the Image dialog box.
2
Click Save Image button to save images to the user-defined folder. (saved as a raw file.)
[TBD] Screen Shot
3
Close the Image dialog box, and open the Calibration dialog box.
4
Choose Defect Calibration as follows and click Manual Defect Correction button.
5
Move to the folder where the raw file is saved and select the file.
[TBD] Screen Shot