User's Manual

Rev.1.0
Page 105 of 151 VW40-153-006
VIVIX-S 1717N User Manual
6.5 Direct Calibration
6.5.1 Prepration
Precheck the state of X-ray generator and console.
Precheck the state of X-ray tube.
It is recommended you to check if X-ray dose value of the generator is exact by using the
device like a dose meter.
Make sure to preheat the detector for 30 minutes before starting calibration. The result of
measurement can be incorrect if the detector is not preheated sufficiently.
6.5.2 Offset Calibration
1
Set the number of Stage as 5 in the Offset Calibration area.
2
Click Run Offset Calibration button and progress the Offset calibration.