6.2 Detector Configuration
From the Detector Configuration dialog, you can configure the detector and check the images before or
after doing calibration.
[TBD] – Screen Shot
Function List
Item
Description
Detector Direction Compensation
Sets the displayed direction of image.
Effective Area
Sets effective area of an image.
Auto Offset Refresh Setting
Sets conditions of automatic Offset Refresh.
Time Interval (min.)
Checks cycle of temperature change
Temperature Interval (℃)
Difference of over-temperature.
Number of shot
Number of Offset Refresh.
Edge Masking (On / Off)
Sets outside of the effective area to specific values.
OSF
Chooses whether to use OSF or not.
Button
Item
Description
Pan
Moves an image to the desired location.
Zoom
Extends or constract an image.
W/L
Adjusts window level of an image.
Fit
Displays an image by adjusting it to the screen.
E.A.
Sets the effective area of an image with a mouse.
ROI
Sets the image area of interest.
Effective Area On / Off
Displays effective area of an image only.
You can only change the displayed direction of an image from Detector Direction
Compensation. The direction of an original image cannot be changed.
Effective Area can be changed within the effective range which has been configured at
the time of the first shipment of detector.
Once Use offset refresh is used, the offset refresh is processed automatically as follows.
Checks temperature difference of the detector between current temperature and the
previous one at the time of offset refresh in every setting time from Time Interval.
If the temperature difference is more than the one set in Temperature Interval, the
offset refresh is carried out a number of times set in Number of shot.