VIVIX-S 1717N User Manual
6.1.4 Gain Calibration Dialogue
Vieworks provides the Gain data stored in the detector. However, you should download
the data as the calibration cannot be performed in the detector automatically.
[TBD] Screen Shot
Gain Calibration
Item
Description
Target Value
Shows the target value of Gain calibration. (Recommended : 8000 or higher)
Current Value
Shows value after the exposure when processing the manual Gain calibration.
Stage
Selects the number of manual Gain calibration and shows each calibration stage.
Get
Starts the manual Gain calibration.
Cancel
Cancels the manual Gain calibration.
Load Gain
Loads the Gain calibration data located in the Gain path.
Upload Gain
Uploads the Gain calibration data to the detector for using portable mode.
Download Gain
Downloads the Gain calibration data stored in the detector.
Gain Path
Shows the path of applied Gain data.
This Gain data is provided in case the Gain calibration cannot be performed. You should
do the Gain calibration in person as the detector condition can be different by the
operating method or use environment.