VIVIX-S 1717N User Manual
6.1.3 Defect Calibration Dialogue
Vieworks provides the Defect Map data stored in the detector. However, you should
download the data as the calibration cannot be performed in the detector automatically.
Defect Map
Item
Description
Load Defect Map
Loads the Defect Map calibration data located in the defect map path.
Download Defect Map
Downloads the Defect Map calibration data stored in the detector.
Manual Defect Correction
Calibrates the defect manually.
Defect Map Path
Shows the path of applied Defect Map data.
This Defect Map data has been generated through stringent test in the production stages
of detector. However, new defect may be newly formed while using the detector. In this
case, make sure to calibrate defects in person and generate a new defect map.