VIVIX-S 1717N User Manual
To acquire the optimized images, carry out the Offset calibration and Gain calibration
again after changing Gain Type.
6.1.2 Offset Calibration Dialogue
Vieworks provides the Defect Map data stored in the detector. However, we recommend
you to carry out the Offset calibration and generate the Offset data by yourself as the
detector condition can be different by the operating method or use environment.
Offset Calibration
Item
Description
Current value
Shows value when performing the manual Offset calibration.
Stage
Selects the performance frequency of manual Offset calibration and
shows each calibration stage.
Run Offset Calibration
Performs the manual Offset calibration.
Download Offset
Downloads the Offset data stored in the detector.
Offset path
Shows the path of the applied Offset data.