VIVIX-S 1717N User Manual
5.4 Changing the Wireless Setting
5.4.1 Switching to the Detector AP Mode
You can change the detector mode as AP by the two ways as follows.
Convert to the Detector AP mode by using the Detector AP button
Press the AP button on the detector for 5 seconds to set the Detector AP mode.
The LED of detector AP blinks in a blue color while the mode is being switched, and turns on blue
after the mode is compeleted to be converted.
This method can be used only when the detector is under the wireless communication
status without connecting a tether interface cable.
Convert to the Detector AP mode from VIVIX Setup
Choose the AP option as On or Off from the Detector Configuration dialog in VIVIX Setup program.
5.4.2 Synchronizing the Wireless Setting
Synchronize the wireless setting with a tether interface cable as follows.
1
Connect the detector and SCU with a tether interface cable.
2
Press the detector AP button for 5 seconds after the detector is turned on.
3
The LED of detector AP blinks while processing synchronization.
4
The sync information is saved automatically after it is transmitted to the detector where SSID and KEY
of SCU are connected.
5
The detector is switched to the wireless communication mode.