VIVIX-S 1717N User Manual
Image
Checks and sets the time limit of image transmission as well as decides whether to use the Preview item.
Item
Description
Time (sec.)
The limited time of completing image transmission.
Preview (Enable / Disable)
Sets whether to use the wireless communication way for sending preview
images.
After starting image transmission, the detector ignores the information of image re-
transmission request if the following conditions are fulfilled.
If the time limit of image transmission does not exceed.
If the image is not transmitted completely.
Power Mode
Checks and sets the management information about the power of detector.
Item
Description
Sleep
Decides whether to use the sleep mode function of the detector. (On / Off)
Sleep after (min.)
If the detector is not used for the specific setting time, it is turned to the sleep
mode. This mode activates only when the sleep mode is set. (10 / 15 / 20 / 25 /
30 min.)
Shut Down
Sets whether to use the shut down function in the detector. (On / Off)
Shut Down after (min.)
The power of detector is off if it is not used within the setting time. This menu
is activated while the Shut Down function is used. (30 / 60 / 90 / 120 min.)
Power Control
Sets standards of power supply to the detector.
by Detector
The detector is operated by power from SCU and battery.
If the power from SCU and battery is blocked, the detector is turned off.
by SCU
The detector is operated by power from SCU with connecting a tether cable. If
the power from SCU is blocked, the detector is turned off.
You can prevent unnecessary battery consumption by using the Sleep function.
When SCU supplies power to the detector with connecting a tether interface cable, the
Shutdown function cannot be operated.
If Power Off is set to by Detector and the tether interface cable is disconnected, you can
keep using the detector by the power of battery. In this case, you can turn off the
detector by pressing the power button for 3 seconds.
If a tether interface cable is disconnected while Power Off is set to by SCU, the detector
is turned off since the battery cannot supply power to the detector.
Wireless network setting should be done by an engineer who understands the wireless
communication and its related technique. Unless the network is set properly, a
communication error would occur or the image quality would be affected.