User's Manual

Rev.1.0
Page 26 of 151 VW40-153-006
VIVIX-S 1717N User Manual
2.3 VIVIX-S 1717N Detector
VIVIX-S 1717N is designed to acquire digital images by collecting x-ray signals and sereval conversion
processes. You can use the acquired image (17 x 17 film size) diversely depending on the purpose of use.
VIVIX-S 1717N is one of the VIVIX detector models produced by Vieworks.
2.3.1 Specifications
Item
Specifications
Model
FXRD-1717NA (CsI)
FXRD-1717NAW (CsI)
FXRD-1717NB (Gadox)
FXRD-1717NBW (Gadox)
Image Sensor
TFT: a-Si (Amorphous Silicon)
X-ray Scintillator Type
FXRD-1717NA(W) : Csl: TI (Thallium doped Caesuim Iodide)
FXRD-1717NB(W) : Gd2O2S:Tb (Gadolinium oxysulfide)
Pixel Pitch
0.14 (140)
Field of View
17 x 17
Active Area (H x V)
430.08 × 430.08
Active Array
3072 x 3072 pixels
Effective Area
428.4 x 428.4
Effective Array
3060 x 3060 Pixels
Grayscale
16 bit
Spatial Resolution
Min. 3.5 lp/
Image Acquisition Time (Wired)
1.5 sec.
Image Acquisition Time (Wireless)
3 sec.
Recommended Cycle Time
15 sec.
X-ray Synchronization Control
AED (Auto Exposure Detection)
DR Trigger (External line trigger)
Rated Power Supply
DC +24V, Max. 1.0A
Wired: Powered by SCU with a tether interface cable.
Wireless: Powered by a battery pack (3,100 h X 2)
Power Consumption
Max. 24W
Dimensions (H × W × D)
460 × 460 × 15.5
Weight
4.2kg (Wired)
4.5kg (Wireless / including a battery pack)
Image Transfer
Wired: Gigabit Ethernet (1000BASE-T) via PoE (Power over
Ethernet)
Wireless: IEEE802.11n
Data Transmission Rate (Wired)
Max. 1Gbps
Data Transmission Rate (Wireless)
Max. 300Mbps (MIMO 2X2)