VIVIX-S 1012N User Manual
6.1.3 Defect Calibration Dialogue
Vieworks provides the Defect Map data stored in the detector. However, you should
download the data as the calibration cannot be performed in the detector automatically.
This Defect Map data has been generated through stringent test in the production stages
of detector. However, new defect may be newly formed while using the detector. In this
case, make sure to calibrate defects in person and generate a new defect map.
Defect Map
Item
Description
Load Defect Map
Loads the Defect Map calibration data
located in the defect map path.
Download Defect Map
Downloads the Defect Map calibration
data stored in the detector.
Manual
Defect Correction
Calibrates the defect manually.
Defect Map Path
Shows the
path of applied Defect Map data.