6.1.2 Offset Calibration Dialogue
Vieworks provides the Defect Map data stored in the detector. However, we recommend
you to carry out the Offset calibration by yourself as the detector condition can be
different by the operating method or use environment.
Offset Calibration
Item
Description
Current value
Shows value
when performing the manual Offset calibration.
S
tage
Select
s the performance frequency of manual Offset calibration and
shows each
calibration stage.
Run Pre
-offset Calibration
Performs the manual O
ffset calibration.
Load Pre
-offset
Downloads the O
ffset data stored in the detector.
O
ffset Path
Shows the
path of the applied Offset data.