User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information

Rev.1.0
Page 92 of 151 VW40-152-009
VIVIX-S 1012N User Manual
5.4 Changing the Wireless Setting
5.4.1 Switching to the Detector AP Mode
You can change the detector mode as AP by following the two ways.
Choose the AP option as On or Off from the Detector Configuration dialog in VIVIX Setup program.
Press the AP button on the detector for 5 seconds to set the Detector AP mode. The LED of detector
AP blinks in a blue color while the mode is being switched, and turns on blue after the mode is
compeleted to be set.
5.4.2 Synchronizing the Wireless Setting
Synchronize the wireless setting with a tether interface cable as follows.
1
Connect the detector and SCU with a tether interface cable.
2
Press the detector AP button for 5 seconds after the detector is turned on.
3
The LED of detector AP blinks while processing synchronization.
4
The sync information is saved automatically after it is transmitted to the detector where SSID and KEY
of SCU are connected.
5
The detector is switched to the wireless communication mode.