User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information

Rev.1.0
Page 47 of 151 VW40-152-009
VIVIX-S 1012N User Manual
3.1 Detector Connection Methods
You can organize wired or wireless connection between the VIVIX-S 1012N detector and SCU. You can also
choose other connection ways suitable for the use environment in case of need.
3.1.1 Wireless Connection
The images and data are transmitted between the VIVIX-S 1012N detector and SCU by wireless
connection.
A battry pack should be installed in the detector to use it under the wireless configuration.
We recommend you to use the wireless connection method for the following cases.
When using the unfixed detector.
When the wired communication is in trouble by the interruption of cables.
The wireless connection is more freely to operate than the wired connection.
Use the wireless connection with a laptop computer to enhance mobility.
Be sure to operate the wireless commnunication between the detector and SCU within a
maximum of 8m.
Use of multiple WLAN devices within the same frequency band may interfere with each
wireless communication and cause a decline in transmission speed.
Do not cover or block the wireless module of the detector. Also, do not put any shielding
materials between the detector and SCU. Otherwise, the transmission speed or operable
distance may be reduced.