User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information

Rev.1.0
Page 131 of 151 VW40-152-009
VIVIX-S 1012N User Manual
2
Check the connection between DC power cable and SCU mini.
3
Tur n off the power switch and turn on again and then check the fan or status of back side.
4
Replace another SCU and check the result of it
5
Replace the fuse of standard SCU. (refer to Change fuse of SCU)
6
Replace corresponding devices.
8.1.3 Communication Test is failed
Symptom
Transmission error is occurred, failure of communication test is occurred.
Expected Causes
Network connection problem
Network setting problem
PC environment setting problem
Wireless environment environment problem
Devices failure
Solutions
1
Check the connection of network cable between Workstation and SCU.
2
Check if the accurate network cable is used or not. (CAT 5E or 6)
3
Set the network information of Workstation, SCU and detector again.
4
Check whole workstation environment again such as firewall setting and release the power save mode.
5
Check surrounding wireless communication environment.
6
Boot up detector and SCU again by processing initialization. (Refer to <7.4 Product Initialization>.)
7
Replace other SCU and Detectors and check the result of it
8
Replace corresponding devices.
8.1.4 The Active LED and Data LED of the Detector are blinking
Symtom
The active LED and data LED is blinking when power LED is ON.
Expected Causes
Detector registeration error
Data transmission error
Solutions
1
Turn on SCU again
2
Check the network cable connection
3
Check the workstation environment and network information again
4
Check if the surrounding wireless communication is good
5
Check cable connection again when connected with tether interface cabel
6
Replace other devices and check the result of it