User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information

Rev.1.0
Page 121 of 151 VW40-152-009
VIVIX-S 1012N User Manual
Self Diagnosis
You can perform self diagnosis of the detector or SCU from the Diagnosis window in the VIVIX Setup
program.
1
Choose a desired item to diagnose.
2
Click Execute Test button located at the bottom of the Test Case window.
3
Check the progress and result of diagnosis for each item in the Result window.
4
Click Save to File to save the diagnosis result as a file when the self diagnosis is completed.
Self diagnosis and resolution can be conducted by a user or a service engineer.
Sensitivity and calibration should be conducted by an authorized service engineer
granted
by Vieworks.
Self-Diagnosis Items of Detector
Category
Item
Form
Expected problem
Action
Voltage
DDC
Decision
Poor condition of a tether
interface cable.
Poor power supply to the wired
operation mode.
Change a tether cable.
Contact a
service engineer.
MAIN
Decision
Poor power supply to the
processor.
Contact a
service engineer.
SIGNAL
Decision
Poor power supply to
FPGA
Contact a service engineer.
Battery
Detection
Decision
Unattached battery
Defective ciruit connected to
a battery pack.
Check if a battery is inserted.
Contact a service engineer.