User's Manual
Table Of Contents
- Instruction
- Product
- System Configuration
- Installation
- Setting
- Calibration
- Diagnosis, Inspection and Maintenance
- Troubleshooting
- Regulatory Information
- Medical Equipment Safety Standards
- Radio Frequency Compliance Information
- Labels and Symbols
- Guidance and Manufacturer’s Declaration for EMC
- Electromagnetic Emissions
- Electromagnetic Immunity
- Electrostatic Discharge (ESD) IEC 61000-4-2
- Electrical Fast Transient/Burst IEC 61000-4-4
- Surge IEC 61000-4-5
- Voltage Dips, Short Interruptions/Voltage Variations on Power Supply Input Lines IEC 61000-4-11
- Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8
- Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3
- Information

Rev.1.0
Page 120 of 151 VW40-152-009
VIVIX-S 1012N User Manual
7.2 Product Inspection
To use products safely, make sure to check the products before use. If problems occur
during inspection or the product is impossible to repair, consult the sales representative
in
Vieworks or a relevant engineer.
7.2.1 Daily Inspection
Before or after using the detector and other surrounding devices, check below items daily.
Item
Description
Detector
Ensure that there are no loose screws or breaks.
Ensure that there is no dust or foreign matter on the battery bay connector.
Ensure that there are no breaks or short-circuits in the battery bay connector.
SCU
Check if the antenna is damaged.
Cable
Ensure that cables are not damaged and cable jackets are not torn.
Ensure that the power cord plugs are securely connected to both AC inlet and AC outlet
of the equipment.
7.2.2 Performance Inspection
Check the detector and other devices periodically as follows.
Item
Period
Description
Self
-Diagnosis
Half
-yearly
Conduct Self-Diagnosis of the VIVIX Setup program for the internal
devices of the detector and check the status.
Resolution
Half
-yearly
Check the resolution of the detector through resolution chart or using a
phantom.
Sensitivity
Half
-yearly
Evaluate the characteristic of the detector through checking gray value of
the images made by X-ray dose amount reaching to the surface of the
detector.
Calibration
Half
-yearly
Updating calibration data. (Offset Gain Defect)
Proceed to calibrate when X-ray Generator, Tube, Collimator or exposure
environment are changed.
Self-diagnosis and resolution can be conducted by a user or a service engineer.
Sensitivity and calibration should be conducted by an authorized service engineer who
Vieworks grants.