VIVIX-S 1012N User Manual
8
Input 15% for Threshold, and 100 for Offset at the bottom of the Pixel List screen.
9
Click Search button at the bottom of the screen. The coordinates of defect pixels and Gray level values
are displayed on Pixel list.
10
While pressing the Ctrl or Shift key, select the coordinate of defect pixels to be added to Defect Map.
11
Click Add button to add the selected items to Pixel List in Manual Defecct Detection.
12
Check if the selected defect pixels are added to the pixel list on the left.
13
Click Save and Close button at the left bottom.