User's Manual

Table Of Contents
Rev.1.0
Page 103 of 151 VW40-152-009
VIVIX-S 1012N User Manual
6.4 Calibrating by Loading the Calibration Data
Vieworks provides the calibration data stored in the detector. Although the Offset
calibration is performed automatically, you need to download both Defect and Gain
calibration data since their calibration cannot be performed in the detector automatically.
6.4.1 Preparing Calibration Data
Among the three types of calibration data stored in the detector, download the Defact Map and Gain
calibration data to local HDD.
It is not necessary to download the Offset calibration data since the data is used in the detector.
File
Description
Offset
Detector Offset data
Gain
Calibration
data for image sensitivity
Defect Map
Defect calibration map data
6.4.2 Loading Defect Calibration Data
1
Select Defect Calibration from the Step list.
2
Click Download Defect Map button to assign the path of local HDD, and download the Defect Map
calibration data.
3
Click Load Defect Map button to load the calibration data.
The calibration menu status is displayed as Done when the Defect calibration data is
loaded completely.