User Guide
Table Of Contents
- Introduction
- Codes, Standards, and Certifications
- PureFlow Press Fittings
- PureFlow Crimp Fittings
- PureFlow Tools
- Product Instructions
- Tubing Overview
- Design Considerations
- ManaBloc Systems
- Installing PureFlow Systems
- Storage Recommendations
- Handling Viega PureFlow PEX Tubing
- Removing PureFlow Connections
- Repairs
- Freezing
- Water Heaters
- Heaters, Flues, Vents, and Recessed Lights
- Fastening the PureFlow System
- Supporting PureFlow PEX Tubing
- Installation Considerations
- Noise and Water Hammer
- Shower Valves
- Electrical Grounding
- Expanding Foams
- Fire Stop Compounds
- Tubing Expansion
- Calculating Expansion Loops and Offsets
- Domestic Hot Water Circulating Systems
- Installing Viega PureFlow Fittings
- PureFlow PEX Tubing Installed in Slab
- Installing PEX Under a Slab
- Installing PEX Below Grade as Service Line
- Pressure Testing PureFlow Systems
- System Disinfection
- Limited Warranty
Pressure Testing PureFlow Systems
56
PG-PF 724998 0320 PureFlow System
Stage 2: Gradually increase system
pressure to 50 psi (over the course of
approximately 1 minute).
■ Wait for 10 minutes before inspecting
pressure gauge.
■ If pressure holds stable, continue to
Stage 3.
■ If pressure falls over the test period,
look for potential leaks with an
approved leak-detect solution. Once
system pressure is completely relieved
and the leak is xed, restart testing
procedure from beginning.
Stage 3: Gradually increase system
pressure to 100 psi (over the course of
approximately 1 minute).
■ Inspect gauge pressure over next
10-minute period to verify pressure has
not decreased more than the allowable
variance noted in the table on the
preceding page.
■ If pressure variance allowed is not
exceeded, the system test has passed.
■ If pressure decreases more than the
allowable variance, the test has failed.
Apply approved leak-detect solution
(see “Leak detection” below) on tting
and/or manifold connections to isolate
the location of the leak. Once system
pressure is completely relieved and the
leak is xed, start testing procedure
from beginning.
Completing the Air Test
Depressurize system.
■ Use of the same safety equipment
applies to these steps.
■ Slowly bleed o any remaining air
pressure from system (approximately
1-minute duration).
■ Notify other trades the test is complete
and the work area is safe to occupy.
■ Document the time and date the test was
completed and the pressure and duration
of the test.
Leak Detection
Viega has identied the leak-detection
solutions listed below as being compatible
for use with Viega PureFlow Press
system components. Consult leak-detect
manufacturer for proper application and
product instructions. To determine the
compatibility of other leak-detection
solutions not listed below, contact Viega
Technical Services prior to their use.
■ megabubble
®
Leak Detector
■ Oatey
®
All Purpose Leak Detector
■ Snoop Liquid Leak Detector
As an alternative leak detection, a mixture
of Original Palmolive Green
™
dishwashing
soap (#46100-46200) or Palmolive Ultra
™
(#356140 or 46128) mixed with potable
water at a ratio of two ounces of soap to
one gallon of water (mix Ultra at a ratio of
1.5 ounces per gallon) may be used.
Note: If the solution does not show a leak
on any of the caps or ttings, isolate the
ManaBloc by turning the valves to the
“OFF” position, repressurize if needed
and apply the same solution to the
ManaBloc manifold components.










