Specifications

Troubleshooting
VTM-200 Installation and Operation Handbook 4-2
Problems, Causes, and Solutions
Table 4-1. VTM-200: Problems, Causes, and Solutions
Problem/Symptom Possible Cause
No video visible on SVGA
monitor.
Output of VTM-200 not connected to SVGA 19 monitor.
SVGA monitor has no power applied.
SVGA monitor cannot lock to
VTM-200 output.
Cable from VTM-200 to monitor not connected properly.
Monitor not capable of 800 x 600 resolution or proper combination of
horizontal and vertical rates.
Incorrect serial digital format being applied.
VTM-200 graticule and display
colors wrong.
Check settings in SETUP menu.
VTM-200 graticules stable but
display is unlocked.
Check to see that input signal is applied to correct connector, and that it
matches the standard selected (NTSC vs. PAL, 525/60 vs. 625/50).
Input signal unterminated.
Input signal is unstable.
External reference selected with no external reference input.
Audio amplitude too small or
too large.
Audio levels and sensitivity not matched.
No audio display or menu
options.
Audio option not installed.
Portions of the display are not
visible (off the VGA screen) or
aspect ratio is incorrect.
Horizontal and vertical position and size on the VGA monitor need
adjustment
If the problem still exists after troubleshooting the VTM-200, see Appendix B,
“Service Support,” for further instructions.