Manual
Table Of Contents
- SmartOTDR Mainframe
- Table of Contents
- About This Guide
- Chapter 1 Safety information
- Chapter 2 Introducing the SmartOTDR
- Chapter 3 Starting up
- Chapter 4 Setting-up the SmartOTDR
- Chapter 5 Optical options
- Chapter 6 Microscope application
- Chapter 7 Connectivity
- Establishing connection
- Remote Control
- Smart Access Anywhere
- Connection modes
- Pre-requisite for using the Smart Access Anywhere Application
- Downloading the VIAVI application on PC
- Downloading the VIAVI application on Tablet/ Smartphone
- Launching the SmartAccess Anywhere application
- Using Remote screen and File Transfer applications
- Working with files and directories on SmartOTDR
- Connection information and settings
- Data Transfer
- VNC
- Smart Access Anywhere
- Stratasync
- Chapter 8 File management
- Chapter 9 Technical specifications
- Chapter 10 Options and accessories
- Chapter 11 Maintenance and Troubleshooting
- Maintenance procedure
- Recycling Information
- Troubleshooting
- General information on warranty
- Index
Chapter 6 Microscope application
Using the microscope
40 User Manual 7SMART102 Rev006
Using the microscope
Running an automated test
1 Launch the microscope application.
2 Connect a fiber to the probe inspection tip.
3 Focus the end face image:
Use the focus control wheel if a P5000i microscope or the autofocus trigger if a
FiberChek microscope.
The focus meter will indicate the focus level.
The focus meter will be green if Advanced Focus = Yes and the focus level is
sufficient for automated analysis.
4 Touch the Test softkey or press the test button on the microscope
The Test softkey will be accessible if the Capture Button setting = Freeze &
Test.
The Freeze softkey will be accessible if the Capture Button setting =
Freeze image.
5 View the end face image and note the test result:
– Pass: green next to result, image outlined in green, green per zone.
– Fail: red next to result, image outlined in red, red next to failing zones.
Figure 28 Exampl
e of the result using the P5000i scope
File Name
Focus Meter
Controls
Fiber End
Face Image