Manual

Table Of Contents
Chapter 9 Technical specifications
Characteristics of the OTDR
102 User Manual 7SMART102 Rev006
Distance displayed takes into account the calibration of the refractive index of the
fiber.
Index adjustable from 1,30000 to 1,70000 in steps of 0,00001
Resolution of display: 1 cm max.
Resolution of cursor: 1 cm max.
Spacing of measurement points: from 4 cm, with up to 256 000 acquisition points.
Accuracy: ±
1m± sampling resolution ±1.10
-5
x distance
(excluding errors of calibration of refractive index of the fiber).
Display span: 0.1 km m to 260 km for single mode
Attenuation measurement
Automatic, manual, 2-point, 5-point, and LSA
Resolution of display: 0,001 dB
Resolution of cursor: 0,001 dB
Linearity: ± 0.04 dB/dB for single mode
Display span: 1.25 dB to 55 dB
Reflectance / ORL Measurement
Resolution of display: 0,01 dB
Accuracy: ± 2 dB
Automatic measurement
Automatic measurement of all the elements of the signal. Slope measurement by
least squares or 2 points of measurement.
Display threshold of faults:
- 0 to 5.99 dB in steps of 0.01 dB for event thresholds
- -11 to -99 dB in steps of 1 dB for the reflectance
- 0.01 to 5.99 dB in steps of 0.01 dB for attenuation
Display of slope and attenuation for a segment of fiber.
Display of the position of a fault and of attenuation.
Display of the reflectance of the fault.
Display of ORL
Manual Measurement
Measurement of slope between the cursors.