20 OLP-35V2/-35SC/-38V2
4 OPERATION
ABSOLUTE POWER LEVEL MODE
Absolute power level mode
The power level is displayed in dBm or Watts (nW, µW, mW).
J Long press [] to toggle dBm/Watts.
NOTE: The context sensitive field shows the selected unit.
Display range
Loss mode
In Loss mode the power level relative to a reference value is
displayed. The reference value can be set by defining the current
power level as the reference value [Abs>Ref].
J Press [] to display relative power level.
Setting the reference level
I Loss mode is selected.
J Press [Abs>Ref].
The current power level is set as the new reference level.
NOTE: The reference level must be stored for each individual
wavelength separately.
Model BN... dBm Watt
OLP-35V2 2333/12 -65 to +10 dBm 1 nW to 10 mW
OLP-35SC 2333/13 -50 to +26 dBm 10 nW to 400 mW
OLP-38V2 2333/14 -65 to +10 dBm 1 nW to 10 mW