OLP-35V2 OLP-35SC OLP-38V2 SmartPocket™ V2 Optical Power Meters User manual BN 2333/98.11 2021.
Please direct all enquiries to your local Viavi sales company. The addresses can be found at: www.viavisolutions.com/en-us/contact-sales-expert The description of additional features of the device can be found at: www.viavisolutions.
CONTENTS CONTENTS I NTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 OLP-35V2/-35SC/-38V2 Optical Power Meters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 User manual update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Symbols used in this user manual . . . . . . . . . . . . . . . . . . . . . 7 SAFETY I NFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 GETTING S TARTED . . . . . . . . . . . . . . . . . . . . . . . .
CONTENTS SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 OLP-35V2/-35SC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 OLP-38V2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 General specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 ORDERING INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . 31 Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 I NTRODUCTION OLP-35V2/-35SC/-38V2 OPTICAL P OWER M ETERS 1 I NTRODUCTION OLP-35V2/-35SC/-38V2 Optical Power Meters The handheld Optical Power Meters • OLP-35V2 (BN 2333/12) • OLP-35SC (BN 2333/13) • OLP-38V2 (BN 2333/14) measure power levels on fiber optical systems. The Test Sets are specially designed for high performance testing of all optical signals and systems, i.e. broadband, PONs, and Gigabit Ethernet.
1 I NTRODUCTION U SER MANUAL UPDATE Differences between the models All the power meters are calibrated at 850 nm, 980 nm, 1310 nm, 1490 nm, 1550 nm, 1577 nm, 1625 nm and 1650 nm. The differences between the devices are the diode types and the maximum permitted power levels: Model BN...
SYMBOLS USED 1 I NTRODUCTION IN THIS USER MANUAL Symbols used in this user manual Various elements are used in this user manual to draw attention to special meanings or important points in the text. Symbols and terms used in warnings The following warnings, symbols and terms are used in this document in compliance with the American National Standard ANSI Z535.6-2011: NOTICE Follow the instructions carefully to avoid damage to or destruction of the instrument.
1 I NTRODUCTION SYMBOLS USED IN THIS USER MANUAL Warning format All warnings have the following format: WARNING Type and source of danger Consequences of ignoring the warning J Action needed to avoid danger. The following character formats are used in this user manual: I Requirement This requirement must be met first; e.g. I The device is switched on. J 1. 2. Italics Instruction Follow the instructions given (the numbers indicate the order in which the instructions should be followed); e.g.
2 SAFETY I NFORMATION SYMBOLS USED IN THIS USER MANUAL 2 SAFETY I NFORMATION J All safety information for your device can be found in the printed booklett “ Safety, Disposal and Environmental Protection“ provided with your device. J Carefully read and follow all instructions given there. J The booklett “ Safety, Disposal and Environmental Protection“ is attached to this pdf. You can open it from the attachment window or by klicking the thumbnail on the left.
3 GETTING S TARTED U NPACKING THE DEVICE 3 GETTING STARTED Unpacking the device Packing material We suggest that you keep the original packing material. It is designed for reuse (unless it is damaged during shipping). Using the original packing material ensures that the device is properly protected during shipping. Checking the package contents Your device is shipped with the following accessories: • 2.
3 GETTING S TARTED D EVICE OVERVIEW Device overview 2 OLP-35SC 2 1 3 4 6 5 OLP-35V2 1 2 3 4 5 6 OLP-35V2/-35SC/-38V2 Test head cover UPP (OLP-35V2/38V2), SC Adapter (OLP-35SC) Display Key pad Representation in the user manual: Context sensitive keys (here left key is selected) [] [MODE] [ G] [ E] Mode/Settings key Save/Results key Power key USB interface For power supply and measurement data downloads and updates.
3 GETTING S TARTED P OWER S UPPLY Keys The key pad contains two types of keys: • Contex t sensitive keys: The functions of these keys depend on the selected mode or menu and is shown in the display above the key. • Function keys: The functions of these keys are always the same and shown on the key itself. Key usage (first and second function levels) The function keys and the context sensitive keys in certain modes have two function levels. A second function level is indicated by two cascading frames.
3 GETTING S TARTED P OWER S UPPLY Battery operation WARNING Dangers in handling batteries Handling batteries may be dangerous. Please note the following safety instructions. J Please note the battery operation safety information in the booklett “Saftey, Disposal and Environmental Protection”. Replacing the batteries J Do not replace individual batteries. Always change both batteries at the same time. J Always use batteries of the same type; i.e. do not mix rechargeable and non-rechargeable batteries.
3 GETTING S TARTED P OWER S UPPLY Protecting the environment Please dispose of any unwanted dry batteries and rechargeable batteries carefully. They should also be removed from the instrument if it is to be scrapped. If facilities in your country exist for collecting waste or for recycling, please make use of them rather than throwing the batteries in the normal trash. You will often be able to return used batteries to the place where you purchase new ones.
4 OPERATION SWITCHING THE DEVICE ON / OFF 4 OPERATION Switching the device on/off The OLP-3xV2 has two battery power modes: Mode Icon Description Permanent ON (PERM) The device is switched on permanently. Automatic OFF (ECON) The device switches off 20 minutes after the last operation. This function is only available when the device is powered from batteries. To switch the device on/off : J Press [ E] to switch the device on/off.
4 OPERATION D ISPLAY ELEMENTS Display elements The following elements can be found in the display. 1 2 3 Display of one value in Auto - mode. Display of two values in Multi- mode. 1 – Status bar Expert/ OptiChek Selected/active mode The mode can be changed via the [MODE] key. Bluetooth® is active Bluetooth connction allows for data transfer via the MobileTech app (for future use). 7:15 Real Time Clock Time can be changed via the settings menu.
4 OPERATION SETTINGS MENU Settings menu The following settings can be changed in the settings menu. J To open the settings menu long press [MENU]. Item Settings Description Hour 1...12 Minute 1...59 Year 2020...2030 Month 01...12 J Press [ ] to change setting: – Press once to change one step at a time. – Hold down the key to increase the step change rate. Day 01...
4 OPERATION SELECTING A WAVELENGTH The wavelengths that can be selected by pressing this key are an extract of those contained in the internal wavelength table. Wavelength 1 Wavelength 2 Wavelength 3 Wavelength 4 Wavelength 5 Wavelength 6 Fig. 2 Wavelength entries Editing the wavelength table The wavelength table supports the definition of up to 6 wavelengths. To edit the wavelength table: I Auto- is switched off. 1. Press [] to select a wavelength. 2. Long press [].
4 OPERATION SELECTING E XPERT /OPTI CHEK MODE Selecting Ex pert/OptiChek mode The OLP-3xV2 provides two operating modes: Expert All functions are available, like selecting a wavelength and loss measurement. Use this mode for in depth evaluation of your test environment. All following described functions are available in Expert mode only. OptiChek This mode is meant for quick and easy test procedures.
4 OPERATION A BSOLUTE POWER LEVEL MODE Absolute power level mode The power level is displayed in dBm or Watts (nW, µW, mW). J Long press [] to toggle dBm/Watts. NOTE: The context sensitive field shows the selected unit. Display range Model BN... dBm Watt OLP-35V2 2333/12 -65 to +10 dBm 1 nW to 10 mW OLP-35SC 2333/13 -50 to +26 dBm 10 nW to 400 mW OLP-38V2 2333/14 -65 to +10 dBm 1 nW to 10 mW Loss mode In Loss mode the power level relative to a reference value is displayed.
4 OPERATION PASS/FAIL MODE Pass/Fail mode An integrated pass/fail analysis feature simplifies standard conformity testing and provides unambiguous measurement result presentation. The OLP-3xV2 provides two Pass/Fail modes: • PF Abs: Pass/Fail indication based on a absolute power threshold • PF Loss: Pass/Fail indication based on a loss threshold To select a Pass/Fail mode: J Press [] until PF Abs or PF Loss is displayed.
4 OPERATION D ISPLAYING MODULATED SIGNALS Displaying modulated signals The OLP-3xV2 automatically detects the modulation frequency of light signals modulated at the fixed frequencies (270 Hz, 330 Hz, 1 kHz, 2 kHz) and CW (constant wave). The detected frequency is shown in the lower center display pane. NOTE: The OLP-3xV2 can automatically detect the modulation frequency only if no Auto- was detected (see “Enabling AutoLambda mode” on page 22).
4 OPERATION M ULTI -L AMBDA MODE Display in Auto- mode When an Auto- signal is detected (Auto- mode needs to be activated), the wavelength setting switches automatically to the applied wavelength and can't be changed manually. Fig. 3 Display in Auto- mode showing one detected wavelength and the power level. Multi-Lambda mode The OLP-3xV2 can receive Multi- signals and display up to three wavelengths simultaneously.
5 M EMORY M ANAGEMENT GENERAL INFORMATION 5 M EMORY M ANAGEMENT General information The OLP-3xV2 allows you to save the measured power level values in a data memory and recall them as required. Up to 1000 results can be stored. NOTE: See also “Specifications” on page 29 for additional data management tools. Storing measurements J Press [ G] to save the current result. The result is saved when "Saving ..." appears on the display and below the name of the currently saved data.
5 M EMORY M ANAGEMENT D ELETING MEASUREMENTS 3. Press [ ] to open the highlighted entry. The selected measurement data is displayed. 4. Press [ ] to return to the list or, press []/[] to show next/previous result or press [MODE] to exit. Deleting measurements I The device shows the list of saved measurements. 1. If you want to delete a single entry, use []/[] to select it. 2. Long press [ ] to open the sub menu. 3. Press [ ] to delete the selected entry. or press [ ] to delete all results.
6 DATA E XPORT AND F IRMWARE U PDATE USB AND S MART R EPORTER 6 DATA E XPORT AND F IRMWARE U PDATE The USB interface or the Bluetooth® interface (OLP-3x only) can be used for data export and firmware update. USB and SmartReporter When using the USB interface the SmartReporter allows you to easily transfer stored measurement data to a PC and to update the firmware. The SmartReporter reporting tool is offered by Viavi for free and can be download from: https://updatemyunit.net/ > Application Software.
7 M AINTENANCE CLEANING THE TEST PORT 7 M AINTENANCE WARNING Invisible laser radiation Maintenance or cleaning of the instrument while it is connected or operating may damage the instrument or injure you. J Make sure that the instrument is switched off and disconnected from all power sources and optical radiation sources before maintenance or cleaning. J Do not open the instrument for maintenance or service. Service shall be performed by Viavi trained personnel only.
7 M AINTENANCE CLEANING THE INSTRUMENT Cleaning the instrument If the device gets dirty through use, you can clean it using a soft cloth moistened with a mild solution of detergent. NOTICE Water and cleaning fluids The instrument may be damaged or destroyed if water or cleaning fluids penetrate it. J Make sure that water or cleaning fluids do not get inside the instrument.
8 SPECIFICATIONS OLP-35V2/-35SC 8 SPECIFICATIONS OLP-35V2/-35SC Adjustable wavelength range Calibrated wavelengths Photo diode Fiber type Display range Resolution Max. power level Intrinsic uncertainty1) Linearity2) Overall measurement uncertainty2) 850 nm, 980 nm 1310 nm, 1550 nm 1490 nm, 1625 nm 1650 nm 800 to 1650 nm (1 nm incr.) 850 nm, 980 nm, 1310 nm, 1490 nm, 1550 nm, 1577 nm, 1625 nm, 1650 nm InGaAs 9/125 to 50/125 -65 to +10 dBm 0.01 dB, 0.001 µW +16 dBm Expert mode: 0.2 dB (5%) OptiChek: 0.
8 SPECIFICATIONS GENERAL SPECIFICATIONS General specifications Wavelength detection1)2) Modulation detection2) Optical adapter system OLS-35V2/-38V2 OLS-35SC Automatic switching and displaying wavelengths CW, 270 Hz, 330 Hz, 1 kHz, 2 kHz UPP 2.5 mm (1.25 mm optional) Fixed SC both suitable for PC and APC systems 1) Only in conjunction with Viavi OLS-3x Optical Light Sources.
9 ORDERING INFORMATION D EVICES 9 ORDERING INFORMATION Devices OLP-35V2 Optical Power Meter InGaAs OLP-35V2 OLP-35SC Optical Power Meter with fixed SC Adapter InGaAs OLP-35SC OLP-38V2 Optical Power Meter, High Power InGaAs (coated) OLP-38V2 Calibration report OLP-35V2, OLP-35SC, OLP-38V2 BN 2333/90.03 Accessories Cleaning materials, power supplies OCK-10 Optical cleaning kit Cleaning tape for optical connectors Spare optical cleaning tape NiMH rechargeable batteries, Mignon AA, 1.
10 P RODUCT R EGULATORY COMPLIANCE A CCESSORIES 10 P RODUCT R EGULATORY COMPLIANCE J All information about the product regulatory compliance can be found in the printed booklett “ Safety, Disposal and Environmental Protection“ provided with your device.
10 P RODUCT R EGULATORY COMPLIANCE A CCESSORIES OLP-35V2/-35SC/-38V2 33