OLP-37XV2 21
PASS/FAIL MODE
4 OPERATION
Pass/Fail mode
An integrated pass/fail analysis feature simplifies standard
conformity testing and provides unambiguous measurement
result presentation.
The OLP-37XV2 provides two Pass/Fail modes:
PF Abs: Pass/Fail indication based on a absolute power
threshold
PF Loss: Pass/Fail indication based on a loss threshold
To select a Pass/Fail mode:
J Press [] until PF Abs or PF Loss is displayed.
NOTE: When PF Abs or PF Loss is selected the two modes can also be
toggled by pressing [].
Setting the fail threshold
Setting the fail threshold is identical for both Pass/Fail modes.
I PF Abs or PF Loss mode is selected. Press [] to toggle
between both modes.
1. Press [] to select a wavelength (see also page 18).
2. Long press [].
The Set Fail Threshold screen opens.
3. Use [+]/[] to change the threshold.
4. Press [
] to save the new threshold.
NOTE: When Dual mode is active, the fail threshold can not be set.
Select G-PON or XGS-PON to set threshhold individually.