MP- and MP-
Specifications for Test Tools
USB plug Run button Patch-cord input
Power LED
Dust cover
MP-60 MP-80
Photo Diode InGaAs InGaAs (coated)
Spectral Range Broadband (800…1650 nm)
Wavelength Setting 800…1650 nm, step size 1 nm
Resolution 0.01 dB, 0.001 µW
Measurement Range for Power
Level
-65…+10 dBm -50…+26 dBm
Max. Power Level +16 dBm +27 dBm
Measurement Uncertainty ± 0.2 dB (± 5%) at ref. conditions
Number of Calibrated
Wavelengths
8
(850, 980, 1310, 1490, 1550, 1577, 1625, 1650 nm)
Tone Detection 270 Hz, 330 Hz, 1kHz, 2kHz
Auto functions
1
Auto-λ
General (typical at 25 °C)
Interface
Electrical interfaces 1x USB-A
Power Supply
AC operation via USB port on PC or laptop
Environmental Conditions
EMI/ESD CE compliant
Recommended calibration interval 3 years
Operating temperature -10 to +55 °C (14 to 122 °F)
Storage temperature -20 to +70 °C (-4 to 158 °F)
Dimensions (H x W x D) 86 x 25 x 19 mm (3.4 x 1.0 x 0.8 in)
Weight
14 g (0.03 lbs.)
When used with VIAVI light sources
USB Power Meters