User's Manual
Chapter 16 Test Results
Copper results
SmartClass TPS User’s Guide
May 2014 22035456, Rev 001 137
Table 45 describes the resistance result.
Distance to short
This result measure the distance to the short. A resistance measurement of 4 k or less typically
indicates a short. For values greater than 4 k, this function is disabled.
Table 46 describes the distance to short result.
Leakage
Leakage is a resistance measurement using higher voltage.
Table 47 describes the leakage result.
Opens/Capacitance
You measure opens/capacitance to:
– measure total electrical loop length (includes length of bridged taps).
Table 45 Resistance result specifications (no ambient)
Range () Resolution ()Accuracy
0 to 999 1 2% ± 2.5
1.0K to 9.99K 10 2%
10.0K to 99.9K 100 2%
100.0K to 999K 1K 2%
1.0M to 9.9M 10K 6.5%
10.0M to 100M 100K 10%
Table 46 Distance to short result specifications
Range (feet) Resolution Range (meters) Resolution
0 to 3000 30 feet 0 to 999 9 meters
3000 to 30,000 100 feet 1000 to 10,000 30.5 meters
Table 47 Leakage result specifications (no ambient)
Range () Resolution ()Accuracy
0 to 49.99 1 2% ± 2.5
50 to 999 1 5% ± 2.5
1.0K to 9.99K 10 5%
10.0K to 99.9K 100 5%
100.0K to 999K 1K 5%
1.0M to 9.9M 10K 10%
10.0M to 100M 100K 15%