Instruction Manual
Page 10 of 15
Accessories
The accessories below are approved for use with OX-A series Oxygen Concentrators:
object/part No.
manufacturer/
trademark
type/model
technical data
mark(s) of conformity
1
)
Bubble
Humidifier(200 ml
Capacity)
Hsiner Co,.Ltd
REF7100
ABS,PP
CE 0434
Mask / Tubing
Hsiner Co,.Ltd
30122
PVC
CE 0434
Cannula w/ 2.1m(7 ft)
Oxygen Tubing
(disposable)
Hsiner Co,.Ltd
3020
PVC
CE 0434
Oxygen Tubing
(disposable)
Hsiner Co,.Ltd
30200
PVC
CE 0434
NOTE- The use of certain humidifiers and accessories not specified above may impair the oxygen
concentrator’s performance.
NOTE- If the Service Required light illuminates and the audible signal alarms but the unit is not operating,
there is no power to the unit. Refer to the Minor Troubleshooting chart on page 8 and contact your
LOVEGO Medical provider.
The OX-10A/5A Device is a device within your concentrator that monitors the oxygen produced by your unit.
When the units is turned “ON,” all four light (Power, Normal Oxygen(Oxygen concentration≥82%),Low
Oxygen(82%>Oxygen concentration≥70%), and Service Required(Oxygen concentration<70%) on the front
panel will illuminate briefly. After a few seconds, only the Power and Normal Oxygen lights will remain lit.
The lights on the front panel are defined as follows:
.Green Normal Oxygen light-acceptable oxygen level.(The LED will display “Normal Oxygen”)
.Yellow Low Oxygen light-below an acceptable oxygen level. (The LED will display “Low Oxygen”)
If the oxygen purity falls below the acceptable level, the green Normal Oxygen light will shut off and the yellow
Low Oxygen light will illuminate. Switch to your reserve oxygen system. Refer to the Troubleshooting section
in this guide on page 8 and contact your LOVEGO provider.
As an added safety feature should the oxygen purity continue to drop, an intermittent audible signal will sound.
Contact your LOVEGO provider immediately. Do not attempt any other maintenance.