Instruction Manual
Revision Date: 08/02/02
LabPro Technical Manual 29
201 Archive Operations Command: Allows the calculator to determine
the contents of LabPro’s
FLASH
memory.
401 Analog Output Setup: This command sets up parameters to
control the analog output driver in LabPro.
1998 Set LED Command: Turns LEDs on and off on command.
1999 Sound Command: Specifies length and frequency of LabPro
sounds.
2001 Direct Output to Digital-Out Port: Outputs data to the digital
output port.
Detailed information about each command is given below. Functionality is continually being added to each
command and the feature is denoted by a firmware version where it first appears.
A table listing valid values follows the syntax of each command. Default values appear in
boldface
type.
The syntax of each command is presented in terms of what is required and what is optional. Every
command must be enclosed in curly brackets, “{“ and “}”. The command # and the option parameters are
comma delimited.
Syntax: { command #, [option [,option…]] }
The command # represents the command to be issued to LabPro. The reset command has the simplest
syntax: s{0}.
[option] indicates an optional parameter. In many instances, using one optional parameter requires
additional optional parameters be specified. The channel setup command is a good example. It can take
one, two or five optional parameters.