VERMEIREN DX2 Basic: without LCDscreen DX2 with LCD screen INSTRUCTION MANUAL
DX2 Operator control vD - 2013-04 OPERATOR CONTROL DX2 1. Use The operator control built into your electric wheelchair enables you to control all the driving, steering, braking and operating processes and to control other adjusting motors (lifting column, seat and back adjuster). The wheelchair's electrical unit and electronics are constantly being monitored internally.
DX2 Operator control vD - 2013-04 1.1. DX2 without LCD screen Press the "ON/OFF" button (1). The charge indicator (2), which also indicates the batteries' charge level, lights up for a moment. If all the lights come on, the batteries are sufficiently charged. If the lights do not all come on, you must adjust your driving activities to suit the low capacity of the batteries. If the battery charger display lights do not come on, check the plug connections of the electronic system.
DX2 Operator control vD - 2013-04 Locking your operator control It is possible to lock your operator control so that unauthorized people cannot turn your electric wheelchair on. To lock your wheelchair: While the power is ON, press and hold the ON/OFF button for 4 seconds. The display will turn off immediately. After 4 seconds all LED's and the Lock LED turn on for 1 second to indicate that the DX system is now locked.
DX2 Operator control vD - 2013-04 Using battery charge indicator The battery charge indicator is used to indicate that your wheelchair power is on and provides an estimate of the remaining battery capacity. Any green LED's indicate that your batteries are well charged. If only amber and red LED's are light, the batteries are moderately charged. Recharge your batteries before you make a long trip. If only red LED's are light, the batteries are running out of charge. Recharge as soon as possible.
DX2 Operator control vD - 2013-04 Adjusting the Driving Speed The DX2 system without LCD screen support 5 drive profiles. A drive profile is a specific drive setting (slow or fast, indoors or outdoors) that your therapist or dealer can adjust to suit your personal needs and preference. You can select the active Drive Profile with the Drive Profile Selection button. Press ‘+’ to increase the drive speed/profile. Press ‘-‘ to decrease the drive speed/profile.
DX2 Operator control vD - 2013-04 The Service indicator light The Service indicator LED is dedicated to displaying fault codes. For a list of the fault codes refer to paragraph 2. If the electric wheelchair has exceeded the drive time and needs maintenance, the fault indicator LED turns on continously for 15 seconds every time the system is turned on. Take your wheelchair to a qualified service center for maintenance. Using the lights Headlights: To turn the Headlights on, press the Headlight Button.
DX2 Operator control vD - 2013-04 Operating the seating functions (adjustment seat, backrest, legrests, …). Move the joystick forward to operate the seating function in one direction. Move the joystick backwards to operate the seating function in the opposite direction. The active seating function LED flashes while operating the actuator. If you move the joystick further forward, the seating function moves faster.
DX2 Operator control vD - 2013-04 1.2. DX2 with LCD screen The function buttons (8), (9) and (10) enable functions to be selected that are displayed in the assigned fields on the screen. The status display is constantly visible in the upper part of the display and always shows the batteries' charge level and the current time. The light, indicator and warning light displays only appear when the relevant function is selected.
DX2 Operator control vD - 2013-04 Select the light function using the field selection button (9) and the display will show the light selection menu. To switch the flash function (left or right) on or off, move the joystick in the appropriate direction left or right. To switch back to the driving program, select the drive profile button (8) in drive mode. Switch the light on/off To switch the light on and off, keep the button for the left-hand indicator (3) pressed for longer than 3 seconds.
DX2 Operator control vD - 2013-04 Drive functions In drive mode use the drive profile button (8) to select a higher or lower driving program (drive profile 1-5). These drive programs are programmed in the factory to rise from slow to fast. The drive program you select appears in red text in the middle of the display. The top speed in any drive program can be changed using the menu selection button (10) in the drive program selected.
DX2 Operator control vD - 2013-04 Adjust function Seat inclination Angle of backrest Leg support left In Display Display Menu L CAUTION: Risk of injury or damage Make sure that no objects and/or persons are inside the swinging range of the adjust functions, since this could cause damage and/or injury. The only functions displayed in the menu selection and the centre of the display are those which exist on your wheelchair and have been enabled.
DX2 Operator control vD - 2013-04 Select the program function using the field selection button (9) and the display will show the menu for the special functions. Select left/right with the joystick or the menu selection button (10) until the time function you want appears in the centre of the display. If you want to change the display option for the time, confirm by moving the joystick upwards. The option currently selected will appear in the menu and the centre of the display.
DX2 Operator control vD - 2013-04 Select left/right with the joystick or the menu selection button (10) until the brightness function you want appears in the centre of the display. If you want to change the brightness function, confirm by moving the joystick upwards. Adjust the display's brightness function by moving the joystick left/right or selecting the left/right menu selection button (10).
DX2 Operator control vD - 2013-04 2. Troubleshooting DX2 Basic (without LCD screen) Symbol flickering fast at intervals. The number of flashes per interval provides the information below (the following actions and checks should only be carried out by authorised individuals). DX2 with LCD screen The following diagnoses refer to all the electronics. The flashing code is read from the status display on the DX2 steering and control unit.
DX2 Operator control vD - 2013-04 Number 8 Problem/Fault Battery overcharged Checks 1. 2. 3. 4. 9 Faulty communication between the power module and the steering unit 1. 2. 3. Check that only the charger supplied has been used. Check that the battery charger is working. This fault may also be displayed if an external power source comes in contact with the wheelchair.
Notes .................................................................................................................................................... .................................................................................................................................................... .................................................................................................................................................... ...........................................................
Notes .................................................................................................................................................... .................................................................................................................................................... .................................................................................................................................................... ...........................................................
Germany N.V. Vermeiren N.V. Vermeirenplein 1 / 15 B-2920 Kalmthout Tel: +32(0)3 620 20 20 Fax: +32(0)3 666 48 94 website: www.vermeiren.be e-mail: info@vermeiren.be Vermeiren Deutschland GmbH Wahlerstraße 12 a D-40472 Düsseldorf Tel: +49(0)211 94 27 90 Fax: +49(0)211 65 36 00 website: www.vermeiren.de e-mail: info@vermeiren.de France Austria Vermeiren France S.A. Z. I., 5, Rue d´Ennevelin F-59710 Avelin Tel: +33(0)3 28 55 07 98 Fax: +33(0)3 20 90 28 89 website: www.vermeiren.fr e-mail: info@vermeiren.