Technical data
7.4.2 Commands and Messages
Setting the operating mode of sources A and B (see 6.2)
OPER_IND - setting of the operating mode INDEPEND (Independent Mode)
OPER_TRAC - setting of the operating mode A-B TRAC (Tracking Mode)
OPER_PAR - setting of the operating mode A-B PAR (Parallel Mode)
OPER? - Current operating mode OPER_IND, OPER_TRAC or OPER_PAR is
saved in the output buffer.
Selecting the sources for the following settings
SEL_A - selection of source A
SEL_B - selection of source B
SEL? - Current selection SEL_A or SEL_B sources A and B is saved in the
output buffer.
Setting the function mode of sources A and B (see 6.3)
CONT_CV - the source works off a constant voltage source
CONT_CC - the source works off a constant current source
CONT? - Current selection CONT_CV or CONT_CC of the function mode is
saved in the output buffer.
Voltage setting of the selected source
VSET xxxxx - voltage setting in the range (xxxxx) from Ø.ØØ to 3Ø.ØØ
VSET_MIN - voltage setting on Ø.ØØ V
VSET_MAX - voltage setting on 3Ø.ØØ V
VSET? - Voltage setting V xxxxx is saved in the output buffer.
VOUT? - Measured voltage value V xxxxx is saved in the output buffer.
Note
: When the selected source works off a constant current source and the
command VSET? is transmitted, the displayed voltage represents the value
of voltage limit.
Current setting of the selected source
ISET xxxxx - current setting in the range (xxxxx)
a) from Ø.ØØ1 to 2.3ØØ for the operating mode INDEPEND and
A-B TRAC
b) from Ø.3ØØ to 4.6ØØ for the operating mode A-B PAR
ISET_MIN - current setting to
a) Ø.ØØ1 A for the operating mode INDEPEND and A-B TRAC
b) Ø.3ØØ A for the operating mode A-B PAR