Instructions
OM17-P4000VIS-NIR
8-6
Sample points latitude column
Spectra latitude column
Spectra depth column (probe mode only)
ports.txt Stores ports and baud rates from
previous time system was run.
ref_wavelengths.txt Stores values of the useable
wavelengths for each spectrometer,
which can be configured by the user.
These wavelengths will be
used to filter the data.
Ex. 450.000 950.000 1150.000 2150.000
These values represent the following controls in this order:
Ocean Optics minimum usable wavelength
Ocean Optics maximum usable wavelength
Hamamatsu minimum usable wavelength
Hamamatsu maximum usable wavelength
probe_xform.txt Stores latest transform.
wlc_*.txt Stores Hamamatsu calibration
coefficients these are provided by Veris.
xform_settings Stores settings for creating transforms,
which the user can configure.
Ex. .075
This value represents the maximum allowed RMSE, when comparing the previous
transform to the current transform.