Mapping System - Trouble-shooting Manual

Coulter Electrode Functions-
Each coulter electrode on the implement is part of a pair, and each pair has a distinct
function.
a) Coulters 1 & 6 are the Deep EC receptors. If you are seeing problems only with the
―Deep‖ readings, focus on testing continuity on these two coulter-electrodes.
b) Coulters 2 & 5 are the ―charged‖ coulters that inject the voltage into the soil. If you
are getting no (or intermittent) readings on both the ―Shallow‖ and the ―Deep‖ --
continuity to one of these two coulter-electrodes is likely the cause.
c) Coulters 3 & 4 are the ―Shallow EC‖ receptors. Focus on this pair if you see
problems in the ―Shallow‖ reading.
If the continuity ohm test indicates a problem on a channel, you will need to determine
where the interruption is located. Listed below are detailed instructions on how to
determine exactly where a continuity or isolation problem is located:
A. Testing Cable and Wiring continuity:
1. Once a high resistance reading on a channel is confirmed, determine whether the
problem is in the wiring or in the coulter-electrode. To test all cable and wiring, place
one ohmmeter lead in the Test Box terminal pin for that channel and the other on
the corresponding coulter wire terminal bolt. Grasp sure-seal connector and move
back and forth during this test vibration from rough fields can weaken the contacts
on the sure seal, causing breaks in continuity during operation that might no show
up in a static test. Repeat process on all coulter-electrodes.
Figure 3.2 Testing cable at coulter electrode #3 terminal
2. If you see <2 ohms on all, test the coulter electrodes as explained in B below.
3. If you see a > 2ohms reading on any channel, separate sure-seal connector and
insert one ohmmeter lead in the end of wiring harness and the other lead in the
corresponding terminal on the Test Box. If reading is < 2 ohms at that point, the