User guide
Bolus Wizard warnings
When using the Bolus Wizard feature, the pump may display a HIGH BG, LOW BG, and MAX BOLUS EXCEEDED
warnings.
HIGH BG
If your blood glucose in the ENTER BG screen is above 13.9 mmol/L (250 mg/dL), the Bolus Wizard feature
displays a HIGH BG warning. Read the instructions, then press ACT or ESC to clear the message. You can then
continue programming to deliver your bolus.
LOW BG
If your blood glucose in the ENTER BG screen is below 3.9 mmol/L (70 mg/dL), the Bolus Wizard feature
displays a LOW BG warning. Read the instructions, then press ACT or ESC to clear the message. You can then
continue programming to deliver your bolus.
MAX BOLUS EXCEEDED
Bolus Wizard feature will not deliver more than the limit set for your maximum bolus. If the Bolus Wizard
feature calculates a bolus amount that is larger than your max bolus limit setting, the message, MAX BOLUS
EXCEEDED will appear. If this happens, do these steps:
1 In the MAX BOLUS EXCEEDED screen, press ACT to continue your bolus programming. The EST : MAX screen
appears with the estimated and maximum bolus amounts. Continue to the next step.
If you do not want to continue, press ESC to cancel and the screen will return to the ENTER BG screen.
2 In the EST : MAX screen, press ACT again to continue your bolus programming.
3 The SET BOLUS screen appears with the maximum bolus amount flashing. Enter the bolus amount. This
amount cannot be more than the max bolus amount. Press ACT.
4 The BOLUS DELIVERY screen appears showing the insulin units being delivered.
5 The pump will beep/vibrate after it has completed the insulin delivery.
How to program the Bolus Wizard feature
You need your personal settings from the Bolus Wizard feature settings table to setup the Bolus Wizard
feature. Your Bolus Wizard feature settings are programmed in the EDIT SETTINGS screen.
Main > Bolus > Bolus Setup > Bolus Wizard Setup > Edit Settings