User guide

Filling the cannula
Filling the soft cannula with insulin is required after the infusion set is inserted into your body and the
introducer needle is pulled out. The insulin amounts required to fill the cannula depend on the type of
infusion set you are using. Refer to your infusion set instructions for this information. If you are using an
infusion set with a needle, press ESC to skip this step.
If you do not complete the Fill Cannula steps, your pump triggers a FINISH LOADING alarm. If you need to
skip the Fill Cannula step, press ESC to avoid getting this alarm.
1 After you have filled the infusion set tubing, the FILL CANNULA screen appears. It displays a message to
connect the infusion set to your body. Press ACT to fill the cannula. To skip this step, press ESC until
you see the MAIN MENU screen. If you do not complete this step, the RESERVOIR + SET MENU screen
appears after 15 minutes. After 10 minutes, your pump triggers a FINISH LOADING alarm.
2 Enter the amount for your type of infusion set, then press ACT. To skip this step, press ESC until you see
the MAIN MENU screen. If you do not complete this step, the RESERVOIR + SET MENU screen appears after
one minute. After 10 minutes, your pump triggers a FINISH LOADING alarm.
3 As the cannula starts filling, the FILLING CANNULA screen will count up the units as they are delivered.
The pump will beep or vibrate when the cannula is full.
To see how much insulin was delivered to fill the infusion set:
1 Go to the HISTORY screen.
Main > Reservoir + Set > History
2 Scroll through the list of insulin deliveries. The letter T at the end of the text line indicates that insulin
was delivered to fill the tubing. The letter C at the end of the text line indicates insulin was delivered
to fill the cannula. Exit the menus.
Starting on insulin70