User guide
WARNING: Make sure the infusion set is disconnected from your body before you rewind your pump or
fill the infusion set tubing. Never insert the reservoir into the pump while the tubing is connected to
your body. Doing so could result in an accidental infusion of insulin.
1 If you removed your reservoir and are replacing it, go to the REWIND screen.
Main > Reservoir + Set > Reservoir Setup
2 In the REWIND screen, press ACT to start the rewind process. The REWINDING screen will appear while
the pump rewinds.
3 After the pump rewinds the REWIND COMPLETE screen will appear.
If you are practicing:
a. Do NOT insert the reservoir in your pump. Make sure the red shipping cap is installed in the reservoir
compartment as shipped.
b. Press ACT, then continue with the instructions in the Filling the tubing section in this chapter.
If you are not practicing, continue to the next section to insert the reservoir in your pump.
Inserting the reservoir in your pump
If your reservoir is already inserted in your pump, continue to the next section.
You must do these steps in the order described. If you are practicing, do NOT insert the reservoir in your
pump.
CAUTION: You must rewind your pump before installing a new reservoir. As part of the pump's function,
it calculates the reservoir volume. To ensure correct volume calculations, your pump has been designed
to require a rewind before you insert your reservoir.
1 If you are using the pump for the first time, remove the red shipping cap from the reservoir compartment.
WARNING: Do not insert the reservoir in the pump if you did not rewind. Doing so could result in
inaccurate insulin delivery.
Never insert the reservoir into the pump while the tubing is connected to your body. Doing so could
result in an accidental infusion of insulin.
2 Insert the reservoir into the top of the pump case.
Starting on insulin 65
Chapter 5