User guide

Changing your infusion set
Removing the reservoir
Each time you remove and replace a reservoir in your pump, you have to rewind your pump, and fill the
infusion set with insulin.
1 Remove the entire infusion set from your body.
2 If attached, remove the activity guard.
3 Turn the tubing connector half-turn counter clockwise, then pull the reservoir and connector out from
the pump.
tubing connector
half-turn counter-clockwise
3
4 Safely dispose of the used reservoir and infusion set item in a sharps container.
5 You must now rewind your pump as described in the next section.
Rewinding your pump
Before you continue, make sure the infusion set is NOT connected to your body and the reservoir is NOT in
the pump.
Starting on insulin64