User guide

Starting on insulin
The Paradigm pump is intended for use with U100 insulin.
Prepare your pump for use
Before continuing with the steps in this chapter, we recommend that you watch your pump training CD-ROM
if available, and complete your pump start training.
When you are done practicing and ready to use your pump with insulin, you must make sure the time and
date are correct on your pump. You must also program your settings as instructed by your healthcare
professional.
You will need these items:
Pump
Insulin (U100)
Paradigm reservoir and user guide
Paradigm compatible infusion set and user guide
Filling the reservoir
WARNING: Using cold insulin can cause air bubbles in the reservoir and tubing. If your insulin is stored
in the refrigerator, allow it to reach room temperature before filling your reservoir. When filling the
reservoir, take care to remove air bubbles.
CAUTION: Insulin comes in different vial types, depending on your country. If your vial is different than
the one described in the User Guide, consult with your healthcare professional on how to fill your
reservoir.
Starting on insulin 61
Chapter 5