User guide
• If you want to use the new BG measurement for calibration, select Save and Calibrate. This option
is available when the Auto Calibration feature is Off and your sensor is communicating with the
pump. The BG measurements must be between 2 mmol/L (40 mg/dL) and 22.2 mmol/L (400 mg/dL).
• If the Bolus Wizard is on and the saved BG is below your BG Target range, the BG Saved screen
appears. This screen indicates that your saved BG is below your BG Target set in the Bolus Wizard
feature, and the pump beeps three times. See Setting the BG Targets, on page 83 for more
information.
• If the Bolus Wizard is on, the saved BG is above your BG Target range, and the calculated
correction estimate is more than the preset scroll rate increment, the BG Saved screen appears.
This screen indicates that your saved BG is above your BG Target set in the Bolus Wizard feature,
and the pump beeps three times. Go to the ENTER BG screen of the Bolus Wizard feature to
enter a correction bolus. See Setting the BG Targets, on page 83 for more information.
• If the information is not correct, select Cancel, then press ACT. The CAPTURE EVENT screen
reappears with Enter BG selected. Repeat the procedure and enter the correct information.
Entering insulin injection information
Follow these steps to enter the amount of insulin you use.
1 Go to the ENTER INSULIN screen.
Main > Capture Event > Insulin Marker
The ENTER INSULIN screen flashes with dashes or the insulin amount that you most recently entered.
2 Enter the amount of insulin you used, then press ACT. A message appears asking you if you want to save
this information. The Yes option is selected.
3 Make sure that the amount of insulin shown on the screen is correct.
• If the information is correct, press ACT. The information you entered is saved to the system and can
now be used in reports.
• If the information is not correct, select No, then press ACT. The CAPTURE EVENT menu shows.
Repeat the procedure to enter the correct information.
Basic programming 51
Chapter 4