When you set the scroll rate to 0.05, you can use the following increments to program a bolus:
0.05 unit increments for a bolus of 9.95 units or less
0.10 unit increments for a bolus of 10 units or more
When you set the scroll rate to 0.10, the increments will increase or decrease by 0.10 units during bolus
programming regardless of the bolus value.
To program your scroll rate:
1 Go to the SET SCROLL RATE screen.
Main > Bolus > Bolus Setup > Scroll Rate
2 The scroll rate value appears flashing. Change the value and press ACT.
3 The pump returns to the BOLUS SETUP screen. Your scroll rate is now programmed and ready to use. Exit
the menus.
Example: Scroll rate
You have set your scroll rate to 0.025 units. You need to set your pump to deliver a 2.5 unit bolus. The
bolus values on the SET BOLUS screen will change in 0.025 unit increments until you reach 1.0 unit. After
this the value will change in 0.05 unit increments until you reach 2.5 units in the SET BOLUS screen.
BG Reminder
After you deliver a bolus you may want to check your blood glucose. The BG Reminder is an optional feature
that makes the pump beep or vibrate to remind you to check your blood glucose after a bolus. However, this
reminder is not available after an Easy Bolus. Your pump is set at the factory with the BG Reminder feature
turned off.
When setting a bolus, if the BG Reminder is on, your pump will ask you to set up the BG Reminder Duration.
This sets how long after bolus delivery you will be reminded to check your BG. This time can be from
30 minutes to 5 hours, or NONE. The BG Reminder is not available after an Easy Bolus.
1 Go to the BG REMINDER SETUP screen.
Main > Bolus > Bolus Setup > BG Reminder
2 Select On and press ACT. The BG Reminder is now enabled. Exit the menus.
Now, the next time you program a bolus, your pump will ask you for the amount of time after your bolus
before you want to be reminded to check your blood glucose.
Basic programming48