User guide
Magnetic fields
Do not use pump cases that have a magnetic clasp.
Do not expose your insulin pump to MRI equipment or other devices that generate very strong magnetic fields.
The magnetic fields in the immediate vicinity of these devices can damage the part of the pump’s motor
that regulates insulin delivery, possibly resulting in over-delivery and severe hypoglycemia.
Your pump must be removed and kept outside the room during magnetic resonance imaging (MRI)
procedures.
If your pump is inadvertently exposed to a strong magnetic field, discontinue use and contact your local help
line or representative for further assistance.
X-rays, MRIs and CT scans
If you are going to have an X-ray, CT scan, MRI or other type of exposure to radiation, take off your pump,
sensor, transmitter, meter and remote control before entering a room containing any of these equipment.
The Paradigm pump is designed to withstand common electromagnetic interference, including airport
security systems. Be sure to carry the Emergency Card provided, when you are traveling. This card provides
airport information.
Precautions
Although the pump has multiple safety alarms, it cannot notify you if the set is leaking or the insulin has lost
its potency. It is essential, therefore, that you test your blood glucose levels at least four times per day. If
your blood glucose is out of range, check the pump and the infusion set to ensure that the necessary amount
of insulin is being delivered.
Avoid extreme temperatures
1 Avoid exposure of your pump and remote control to temperatures above 42°C (108°F) or below 1°C
(34°F).
2 Insulin solutions freeze near 0°C (32°F) and degrade at high temperatures. If you are outside in cold
weather, wear your pump close to your body and cover it with warm clothing. If you are in a warm
environment, take measures to keep your pump and insulin cool.
3 Do not steam, sterilize or autoclave your pump or remote control.
Introduction6