User guide
Recommended separation distances between portable and mobile RF communications
equipment and the Paradigm 554/754 pump system
This section provides information on the recommended separation distance between portable and
mobile RF communications equipment and the Paradigm 554/754 pump. The Paradigm 554/754 pump
is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. Paradigm 554/754 pump users can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Paradigm 554/754 pump as recommended below, according to the maximum
output power of the communications equipment.
Separation distance according to the frequency of transmitter (m)
Rated maximum
output power of
transmitter (W)
80MHz to 800MHz
d = 0.35
P
800MHz to 6.0GHz
d = 0.70
P
0.01 0.035 0.07
0.1 0.11 0.11
1 0.35 0.7
10 1.1 2.2
100 3.5 7
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
p is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Pump specifications224