User guide

Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 554/754 pump is intended for use in the electromagnetic environment specified below.
The customer or user of the Paradigm 554/754 pump system should make sure that it is used in such
an environment.
Immunity
Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment Guidance
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption, and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcast cannot be predicted
theoretically with accuracy. To access the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the Paradigm 554/754 pump is used exceeds the applicable RF compliance level above, the Paradigm
554/754 pump should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the Paradigm 554/754 pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.
Pump specifications 223
Chapter 15